64E-5.644. Radiation Surveys for Remote Afterloader and Gamma Stereotactic Radiosurgery Facilities  


Effective on Thursday, February 11, 2010
  • 1(1) 2The licensee shall perform radiation surveys with an operable radiation survey instrument calibrated as provided in Rule 1964E-5.615, 20F.A.C., to ensure that the maximum radiation levels and average radiation levels from the surface of the main source safe with the source(s) in the shielded position 47do not exceed the levels stated in the Sealed Source and Device Registry.

    60(2) 61The licensee shall make the survey specified in subsection 7064E-5.644(1), 71F.A.C., at the installation of a new source and following repairs to the source(s) shielding, the source(s) driving unit, or other electronic or mechanical component that could expose the source, reduce the shielding around the source(s), or compromise the radiation safety of the unit or the source(s).

    118(3) 119A licensee shall retain a record of the radiation surveys required by subsection 13264E-5.644(1), 133F.A.C., for the duration of the license. These records shall include:

    144(a) The date of the measurements;

    150(b) The manufacturer's name, model number and serial number of the treatment unit, source, and instrument used to measure radiation levels;

    171(c) Each dose rate measured around the source while the unit is in the off position and the average of all measurements; and

    194(d) The signature of the RSO or authorized medical physicist who performed the test.

    208Rulemaking Authority 210404.022, 211404.051, 212404.061, 213404.071, 214404.081, 215404.141 FS. 217Law Implemented 219404.022, 220404.051(1), 221(4), (5), (6), (8), (9), (10), (11), 228404.061(2), 229(3), 230404.071(1), 231404.081, 232404.141 FS. 234History–New 8-25-91, Formerly 10D-91.763, Amended 2-11-10.

     

Rulemaking Events:

Historical Versions(1)

Select effective date to view different version.