64E-5.645. Remote Afterloader, Gamma Stereotactic, and Teletherapy Therapy-Related Computer Systems  


Effective on Thursday, December 26, 2013
  • 1The licensee shall perform acceptance testing on the treatment planning system of high, medium, low, pulsed dose-rate remote afterloaders, gamma stereotactic, and teletherapy therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. An example of a nationally recognized body is the American Association of Physicists in Medicine. At a minimum, the acceptance testing must include, as applicable, verification of the following:

    66(1) 67The source-specific input parameters required by the dose calculation algorithm;

    77(2) 78The accuracy of dose, dwell time, and treatment time calculations at representative points;

    91(3) The accuracy of isodose plots and graphic displays;

    100(4) The accuracy of the software used to determine sealed source positions from radiographic images; and

    116(5) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.

    137Rulemaking Authority 139404.051, 140404.061, 141404.071, 142404.081, 143404.141 FS. 145Law Implemented 147404.022, 148404.051(1), 149(4), (5), (6), (8), (9), (10), (11), 156404.061(2), 157(3), 158404.071(1), 159404.081, 160404.141 FS. 162History–New 8-25-91, Amended 1-1-94, Formerly 10D-91.764, Amended 10-8-00, 2-11-10, 12-26-13.

     

Rulemaking Events:

Historical Versions(2)

Select effective date to view different version.