64E-5.6423. Additional Technical Requirements for Mobile Remote Afterloader Units  


Effective on Thursday, February 11, 2010
  • 1(1) A licensee providing mobile remote afterloader service for medical use shall perform the following:

    16(a) Check survey instruments before medical use at each address of use or on each day of use, whichever is more frequent; and

    39(b) Account for all sources before departure from a client’s address of use.

    52(2) In addition to the periodic spot-checks required by Rule 6264E-5.6421, 63F.A.C., a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of the following:

    103(a) Electrical interlocks on treatment area access points;

    111(b) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;

    129(c) Viewing and intercom systems;

    134(d) Applicators, source transfer tubes, and transfer tube-applicator interfaces;

    143(e) Radiation monitors used to indicate room exposures;

    151(f) Source positioning (accuracy); and

    156(g) Radiation monitors used to indicate whether the source has returned to a safe shielded position.

    172(3) In addition to the requirements for checks in subsection 18264E-5.6423(2), 183F.A.C., a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.

    210(4) If the results of the checks required in subsection 22064E-5.6423(2), 221F.A.C., indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.

    257(5) The licensee shall keep a copy of each check for mobile remote afterloader unit required by subsection 27564E-5.6423(2), 276F.A.C., for three years. The records shall include:

    284(a) The date of the check;

    290(b) The manufacturer’s name, model number, and serial number of the remote afterloader unit;

    304(c) Notations accounting for all sources before the licensee departs from a facility;

    317(d) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators, source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and

    352(e) The signature of the individual who performed the check.

    362Rulemaking Authority 364404.022, 365404.051, 366404.061, 367404.071, 368404.081, 369404.141 FS. 371Law Implemented 373404.022, 374404.051(1), 375(4), (5), (6), (8), (9), (10), (11), 382404.061(2), 383(3), 384404.071(1), 385404.081, 386404.141 FS. 388History–New 2-11-10.

     